ECG
• Input dynamic range: ±(0.5~5 mVp)
• Differential input impedance: =10 mO
• Bandwidth:
– 0.05~150 Hz (diagnostic)
– 0.5~40 Hz (monitoring)
– 1~20 Hz (operation)
• CMRR:
– =90 db (diagnostic)
– =105 db (monitoring & operation)
• Sensitivity selection: ×1/4, ×1/2, ×1, ×2, ×4 and Auto
• Sweeping speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
• HR measuring range: 15~350 bpm ±1% or ±2 bpm, whichever is greater
• Pacemaker pulse detection and rejection function
RESP
• Measuring range: 0~120 rpm ±5% or ±2 rpm, whichever is greater
TEMP
• Measuring range: 21.0~50.0 °C ±0.2 °C from 25~45 °C
NIBP
• Technique: oscillometric method
• Typical measurement time: <30 seconds (adult cuff)
• NIBP measuring range SYS:
– 40~275 mmHg (Adult)
– 40~200 mmHg (Pediatric)
– 40~135 mmHg (Neonate)
• NIBP measuring range DIA:
– 10~210 mmHg (Adult)
– 10~150 mmHg (Pediatric)
– 10~95 mmHg (Neonate)
• NIBP measuring range MAP:
– 20~230 mmHg (Adult)
– 20~165 mmHg (Pediatric)
– 20~110 mmHg (Neonate)
• NIBP measuring accuracy:
– Mean difference: ±5 mmHg
– Standard deviation: 8 mmHg
• NIBP measurement mode: Manual, Auto, Stat, Multi-cycle mode
• Auto measuring intervals: 1-480 min
SpO2
• Technique: Dual-wavelength optical method
• Measuring range: 0%~100% ±2% for Spo2 range 70~100%
• PR measuring range: 30~250bpm ±2bpm or ±2%, whichever is greater
• Low perfusion performance: as low as 0.3%.
CO2
• Technique: infrared optical method
• Sampling mode: Sidestream or mainstream
• Measuring range: 0~150 mmHg
• Measuring accuracy:
– 0~40 mmHg ±2 mmHg
– 41~70 mmHg ±5% of reading
– 71~100 mmHg ±8% of reading
– 101~150 mmHg ±10% of reading
• Flow rate: 50 ml/min ±10 ml/min (Sidestream)
Other specifications
• Power supply: AC 100-240 V, 50/60 Hz, 60 VA
• Built-in lithium battery: 11.1V/4400 mah
• Display: 15 inch TFT display
• Alarming method: 3 levels audible-visible alarm
• Networking: Ethernet
• Safety standard: IEC 60601-1