General requirements
• Dimension: 335x225x360 mm
• Weight: 6.5 kg with dual batteries
• Electrical: AC 100-240 VAC, 50/60 Hz
• Rechargeable Lithium-ion batteries: 15.0 V – 3.2 Ah
• Battery Operation Time:
>5 hours, with 1 battery
>10 hours with 2 batteries
• Charge Time: Max. 3 hours
• Capacity: 150 shocks with 1 battery or 300 shocks with 2 batteries at 230J fully charged
• Software: Voice and texts multi language
• Memory: Internal flash for ECG data, events and usage review
Defibrillation
• Waveform: Biphasic Truncated Exponential (BTE), HiCAP technology (Large storage capacitor)
• Impedance: C ompensated, range 25-200 Ohm
• Charge time: 6 sec for 230 J with battery fully charged
AED mode
• Energy: Fixed energy at 200 J
• Protocol: ERC 2015 CPR Guidelines with voice and text prompts
• Shockable rhythms: VF with amplitude > 0.15 mV and VT with rhythm >150 bpm
• Sensitivity: S hockable rhythm -VF > 95%, -VT > 75%
• Specificity: According to ERC 2015,
• Normal sinusoidal rhythm > 99%
• Asystole and other non-shockable rhythms > 99%
Manual mode
• Energy range: 1-230 J (from 1-10 J in 1 J steps; from 10 to 230 J in 10 J steps)
• Commands: Multifunction trim knob
• Charge and shock button in the front panel
• Operating mode: EC G «R» wave synchro or async mode
• Indicators: B attery and main led indicators
Display
• Type: High contrast TFT colour LCD 8” (800×600 dots)
• Information: ECG traces, SpO2, NIBP, EtCO2, TEMP, Heart Rate indicator, filters, alarms and batteries status
• Display mode: Can select 3, 6, 12 ECG leads
Ecg monitoring
• Bandwidth:
0.05 – 120 Hz (-3 db) in diagnostic mode
0.6 – 40 Hz (-3 db) in monitoring mode
• CMRR: > 98 dbv Input impedance: > 50 Mohm
• Input:
Patient cable 5 leads (I, II, III, aVL, aVF, aVR, V) – standard
Patient cable 10 leads (I, II, III, aVL, aVF, aVR, V1…..V6) – optional
• Speed: 5, 10, 25, 50 mm/sec
• Filters: 50/60 Hz, EMG Filter, base line
• HR range: 20-300 bpm (~2%), digital readout on the display
• Alarm: HR max settable 250 bpm; HR min 20 bpm
SpO2 (only 55505-55512)
• SpO2 range: 0-100% HR range: 30-250 bpm
• Accuracy: 70-100% ~2% for adults with finger clip sensor
• Alarm: Adjustable min 50%
Oscillometric NIBP (only 55510-55512)
• Accuracy: Meets ANSI/AAMI SP10-2002, EN 1060-4
• Application: Adult / Paediatric / Neonatal (mmHg)
• Systolic Range: 40-260 40-160 40-130
• Range MAP: 26-220 26-133 26-110
• Diastolic Range: 20-200 20-120 20-100
• PR range: 30-220 bpm ±2% or ±3 bpm
• Operating Modes: Manual, Long Term Automatic, Stat, Service
• Auto Interval Periods: 1,2,3,4,5,10,15,30,60 and 90 minutes
Integrated thermal 3 channels printer
• Paper speed 5, 10, 25, 50 mm/sec Paper width: 58 mm
• Mode: Manual, automatic (12” pre and post shock recording)
External pacemaker (only 55505-55512)
• Type: Rectangular wave
• Operating mode: Fixed on demand
• Pulse rate: 30 to 250 ppm, adj. in steps of 5 ppm
• Impulse duration: 22.5 ms
• Pulse current: 0 to 180 mA adj. in steps of 5mA
• Amplitude: Max 200 V
Mainstream technology Entidal CO2 (optional)
• Suitable for both intubated and non-intubated patients
• Measurement range: 0 to 150mmHg ±2%
• Respiratory rate: 0 to 150 bpm
Temperature
• Probe type: YSI400 compatible
• Range: 0 to 50° C
• Resolution: ±0.1° C
Standard
• IEC 60601-2-4; IEC 60601-1; IEC 60601-1-2; IEC 63204, IEC 62366-1
• CE Mark EEC 93/42: Medical device, Class IIb
• Isolation: ECG type CF, Defibrillation type BF
• Water proof: Class IP54