MONITOR
• Display, Resolution: 12.1“ color TFT. 800×600 pixels
• Size 322 x 250 x 224.8 mm
• Weight Approx.: 4.5 kg
• Trace:
– 6 waveforms: 2xECG, SpO2, RR or EtCO2, 2xIBP
– Sweep speed: 6.25, 12.5, 25, 50 mm/sec
• Indicators:
– Categorized alarms (3 priority levels)
– Visual alarm lamp handle and heart beat tone
– QRS beep & SpO2 pulse beep
– SpO2 pitch tone
– Battery status, external Power LED
• Interfaces: Touch screen, rotary knob
• Connectors:
– DC input connector: 18 VDC, 2.5 A
– Defibrillator Sync. Output:
– signal level: 0 to 5 V pulse
– pulse width: 100 ± 10 ms
– LAN dIgital output for transferring data
– External VGA monitor connection
– DC output 5VDC. 1A Max
• Battery: Rechargeable Li-ion battery, 1.5 hours continuous operation
• Thermal Printer:
– speed: 25, 50 mm/sec,
– paper width: 58 mm
• Data Storage:
– 168 hours trends, 20 cases of 10 sec alarm waveform
– USB memory storage
• Internal language (selectable) GB, FR, IT, ES, DE, CZ, PL, PT, RO, BG, HU, TR, RU, CN, KR
ECG
• Leads type: 3-lead, 5-lead
• Lead selection:
– 3-lead: I, II, III
– 5 lead: I, II, III, aVR, aVL, aVF, V
• ECG waveform:
– 3-lead: 1 channel
– 5-lead: 2/7 channel
• Heart rate range:
– Adult: 15-300 bpm
– Neonate/pediatric: 15-350 bpm
• Heart rate accuracy: ±1 bpm or ±1% wichever is greater
• Sweep speed 6.25, 12.5, 25, 50 mm/sec
• Filter:
– Diagnosis: 0.05~120 Hz
– Monitoring: 0.5~40 Hz
– Moderate: 0.5~25 Hz
– Maximum: 5~25 Hz
• S-T segment detection range: -2.0 to 2.0 mV
• Arrhythmia analysis: ASYSTOLE, VTACH, VFIB, BIGEMINY, ACCVENT, COUPLET, IRREGULAR, PAUSE, PVC, RONT, TRIGMINY, VBRADY, SHORTRUN
• Pacemaker detection mode: Indicator on waveform display (user selectable)
• Protection: Against electrosurgical interference and defibrillation
RESPIRATION
• Method: Thoracic impedance
• Channel selection: RA-LA or RA-LL
• Measurement range: 5-120 breath per minute
• Accuracy: ±1 breath per minute
• Apnea alarm: Yes
• SpO2 Saturation range: 0 to 100%; (70 to 100% ±2% digits)
• Pulse rate range: 0 to 254 bpm ±2 bpm
NIBP SUNTECH
• Method: Oscillometry
• Operation Mode: manual / Automatic / Continuous
• Pressure measurement range: 0-300 mmHg
• Safety pressure:
– Adult: 300 mmHg
– Neonate: 150 mmHg
• Pulse rate: 30-220 bpm
• Blood pressure:
– Adult pressure: 20 to 220 mmHg
– Pediatric pressure: 20-160 mmHg
– Neonate pressure: 20-130 mmHg
• Accuracy:
– The static pressure measurement is within ±3 mmHg
– Meets ANSI/AAMI SP 10 2002
TEMPERATURE
• Measurement range: 10 to 45°C (59 to 113°F) ±0.1°C
• Compatibility YSI SERIES 400 temperature probes
IBP
• Channels 2
• Measurement range: -50 to 300 mmHg
• Accuracy:
– -100 mmHg ±1% of reading
• Pulse rate measurement range: 0 to 300 bpm
• Zero balancing:
– Range: ±200 mmHg
– Accuracy: ±1 mmHg
– Drift: ±1 mmHg of over 24 hours
• Transducer sensitivity: 5 µV/mmHg
• Pulse rate measurement range: 0 to 300 bpm